Avia Pulse D Series

Portable Pulse Oxygen System (PPOS) for therapeutic use by customers with a pre-existing medical condition.

Overview of the Avia Pulse D Series

The Avia Pulse is a thereputic Oxygen Conserving System designed for passengers who have a pre-exisiting medical condition and pre-book oxygen for their flight(s). The advantage of the Avia Pulse Dose is that oxygen is only consumed during that part of the inhalation where the oxygen is of benefit and a reduced amount of oxygen is therefore needed therefore nedded with the consequent weight saving.

The D series has 5 pulse settings and a constant flow setting of 2 litres per minute. The system compromises of a lightweight DOT approved cylinder, available in 3 sizes; 2.0 litre, 2.7 litre and 3.2 litre water capacity with a working pressure of 200 bar.

The conserver gives extremely economical performance, releasing a bolus of oxygen as soon as it registers a small drop of pressure in the passenger's nose, it then closes very quickly, thus conserving oxygen. The conserver performance is best in class.

The regulator is of robust design and incorporates a number of safety features.

The 'cage' is designed to protect the regulator, conserver, pressure gauge, ease handling and present a softer image to the passengers and crew. It is manafactuered in a flame retardant material and has a Contact Memory Button (CMB) fitted at the rear, which stores data on configuration control and traceability of the assembly in accordance with ATA Spec 2000.


A purpose designed flame retardant bag, includes pockets for the operating instructions, cannula and loops for attachment. Cannula assemblies can be supplied with an inline flow indicator, allowing cabin crew to check delivery of oxygen whulst a passenger is sleeping.

Human factors have been incorporated in to the design from the outset and Avia Technique has been incorpoated a number of features following consultation with experienced airline cabin safety specialists and engineers.

Key Features

  • Very simple to use
  • Excellent duration
  • Safety, able to withstand a 3 metre drop test
  • Totally pneumatically operated, no batteries or extenal power required
  • All brass high pressure components
  • Medically proven
  • CMB carries full tracability of the assembly

Cylinder Specification

The table below shows the performance of the Avia Pulse D Series cylinder under each individual setting.

                 
Size O2 Capacity (L) Cylinder Pressure 200 Bar Setting 1 hours Setting 2 hours Setting 3 hours Setting 4 hours Setting 5 hours 2 LPM continious flow hours
2L 400 190 22.6 14.4 10.9 8.8 7.2 3.2
2.7L 540 190 30.5 19.4 14.7 11.9 9.7 4.3
3.2L 640 190 36.2 23.0 17.4 14.1 11.5 5.1
                 
  • Operation

  • 1) Check 02 levels are at full

  • 2) Attach cannula to bottle

  • 3) Select correct flow rate as per customer requirements

  • 4) Turn the isolation valve to fully open

  • 5) Customer to postion cannula on face

  • 6) Ensure customer has oxygen flow by flow indiciator

  • 7) Check customer and gauge every 30 minutes

AS9100 Revision B (Design Approval)
BS EN ISO 9001 - 2008 (Design Approval)
FAR 145
TCCA Transport Canada
US DOT Hydrostatic Testing
EASA Part 21 Subpart G
EASA Part 145